Controlling NOx effectively: SICK supports the optimization of DeNOx plants

Aug 6, 2020

Environmental awareness is growing. As in all major manufacturing industries worldwide, this also applies to the cement industry. The emission of hazardous substances and environmental pollution should be reduced or prevented in any possible way. To effectively support climate protection, to maintain and restore a clean environment are the most important goals of modern cement producers. The motivations are manifold. Be it due to new environmental legislation, pressure from the local population, external organizations and interest groups of the company or self-imposed sustainability obligations.

GM32 on site during preventive check-up
Figure 1: GM32 on site during preventive check-up
GM32 on site during preventive check-up
Figure 1: GM32 on site during preventive check-up

 

NOx emissions and regulations

Nitrogen oxides (NOx) and other nitrogen compounds arising from cement production are considered the main reason for photochemical smog and formation of nitric acid and acid rain. They are formed either by a combination of fuel-based nitrogen with oxygen within the flame or by a combination of atmospheric nitrogen in the combustion air. The two main mechanisms for the production of NOx in a clinker burning process are the reaction of nitrogen in combustion air at high temperatures (thermal NOx) and the combustion of nitrogen containing fuels (fuel NOx). Globally the emission limits for NOx show the same trend towards lower emission standards and stricter penalties for non-complying plants. However, there are still big differences for this pollutant. Most national limits are in the region of 500 to 1,000 mg/Nm3. Some plants in the EU take pioneering roles by having fixed emission limit values of down to 200 mg/Nm³, depending on which kiln process is applied and what type of fuels are being used.

 

Even though primary measures such as flame cooling, installation of low-NOx burners, staged combustion and general process optimizations show significant effects in reducing NOx emissions, secondary measures are seen as vital for NOx abatement and lower emission limit values. In cement production two secondary measures, selective catalytic reduction (SCR) and selective non-catalytic reduction systems (SNCR), have constantly gained in popularity over the last years and are accepted and proven technologies when controlled in a proper way.

 

SCR and SNCR – two technologies, one goal

SCR and SNCR denitrification (DeNOx) plants use ammonia (NH3) for an efficient reduction of NOx to form the harmless nitrogen and water. The main difference between both technologies is the use of a catalyst. The SNCR, i.e. DeNOx without a catalyst, is installed in the riser duct or calciner after the rotary kiln at a temperature range of 900 to 1.000 °C. Depending on the type of the SCR it can be placed in the high dust raw gas stream (e.g. directly after the pre-heater) or before the main stack in the low dust gas stream as a so-called tail-end or low-dust SCR. A SCR consists of a specific number of catalyst layers and operates at gas temperature of around 300 to 350 °C. The reducing agent, typically ammonia water or urea, is injected at the SCR inlet. For an initial reduction of NOx emission values, often a SNCR is sufficient and lower in cost to install than a SCR. An efficient reaction between the NH3 and NOx depends highly on the temperature. At temperatures greater than 1.200 °C, NH3 converts to NOx, at temperatures lower than 800 °C, the ammonia slip increases. Thus, an efficient and well-adjusted control of the ammonia injection is of high importance for the compliance of gas emission limits when using a SNCR. The presence of the catalyst in the SCR allows the operation at lower temperatures and offers a higher stoichiometric ratio, which leads to a lower amount of injected reducing agent and ensures a lower NH3 slip. Thus, very low NH3 and NOx emission limit values can be adhered to.

 

 

Figure 1: Cement diagram to illustrate the production and gas flow and identifying the measuring point
Figure 2: Cement diagram to illustrate the production and gas flow and identifying the measuring point
Figure 1: Cement diagram to illustrate the production and gas flow and identifying the measuring point
Figure 2: Cement diagram to illustrate the production and gas flow and identifying the measuring point

 

The project HeidelbergCement Geseke

With the growing demand for SCR and SNCR units in cement production, the need for a measurement technology that can be used to control these plants efficiently is also increasing. According new emission limits from 01.01.2019, which have been released in (17. BImSchV; Federal emission control act) 200 mg/Nm3 for NOx and 30 mg/Nm3 for NH3 is required. To comply with local emission regulations, HeidelbergCement decided to invest in a SCR plant in addition to the already existing SNCR solution. HeidelbergCement Geseke finished the commissioning of the new high dust SCR plant in March 2019.

 

During planning and implementation of the project, the question of which kind of analyzer could fulfill the requirements for an efficient SCR control was addressed. The device should be placed at the SCR inlet between the ammonia water injection nozzles and the catalyst (see fig. 3).

 

Of high importance were the fast response time of the analyzer for a very efficient control of the ammonia water injection as well as a long maintenance free period under these very challenging conditions. The advantage of the measuring location at the inlet is the simultaneous measurement of NH3 and NO entering the SCR. Here, NH3 is the sum of ammonia evaporation from the raw materials, the ammonia slip of the SNCR and the ammonia water injection for the SCR. As well, the NO originating from the combustion process could be measured. With this measurement and the Continuous Emission Monitoring System (CEMS) at the main stack as backup, a feed forward control would be possible .

 

Within the project-planning phase, HeidelbergCement contacted SICK to provide a solution, which best fits the requirements mentioned above. As the only provider for all major gas measuring principles and technologies alongside with many years of experience, SICK was able to select and provide the best measuring technology for each application. In this case, SICK decided to provide a GM32 in-situ analyzer in order to cope with the very challenging process conditions — high dust, high temperature and vibrations.

 

 

Figure 3: Installation point of GM32 onsite at the SCR inlet at HeidelbergCement Geseke plant
Figure 3: Installation point of GM32 onsite at the SCR inlet at HeidelbergCement Geseke plant
Figure 3: Installation point of GM32 onsite at the SCR inlet at HeidelbergCement Geseke plant
Figure 3: Installation point of GM32 onsite at the SCR inlet at HeidelbergCement Geseke plant

 

GM32 – the innovative in-situ gas analyzer

The in-situ gas analyzer GM32 from SICK measures simultaneously up to four components (NO, NO2, NH3, SO2) plus temperature and pressure directly inside the process gas stream. The direct measurement inside the duct (in-situ) leads to fast measuring results due to a short response time, which is perfectly suitable for process control. The GM32 analyzer unit is equipped with a gas permeable probe (GPP), which is positioned inside the duct (see fig. 4). Using the wavelength-specific light absorption by the gas mixture on the active measuring path, the sender/receiver unit determines the concentration of the gas components present. UV light sent from the sender/receiver unit passes the active measuring path of the GPP probe and is reflected by a triple reflector at the end of the probe. The permeable filter element — the heart of the GPP— keeps all dust outside of the measuring path, where the light passes through, while the gas permeates quickly through the pores ensuring the required fast response time. The GM32 uses the DOAS principle (Differential Optical Absorption Spectroscopy) where the absorption lines of specific gases in a particular wavelength range are evaluated.

 

Neither the filter, nor the rest of the gas analyzer require weekly or monthly checking, cleaning or other high frequent maintenance work.

 

Due to the higher temperature and possible temperature fluctuations at the measuring location, stack movements are possible. With the implemented auto alignment correction, which aligns the light beam continuously during operation, stack movements as well as vibrations can be compensated for. This ensures a stable and reliable measurement.

 

In comparison to many other measuring systems, which require a frequent test gas calibration, the integrated filters for zero and span check (approved according to EN15267) automatically compensate drifts and ensure a correct and accurate measurement. This helps in addition to keep operational expenditures very low.

 

 

Figure 4: GM32 measurement technology: optical path and UV absorption spectra
Figure 4: GM32 measurement technology: optical path and UV absorption spectra
Figure 4: GM32 measurement technology: optical path and UV absorption spectra
Figure 4: GM32 measurement technology: optical path and UV absorption spectra

 

After the test run on the long distance

The final commissioning of the in-situ gas analyzer at HeidelbergCement plant Geseke took place in March 2019 (fig. 3). During the 12-month testing period (March 2019 – March 2020), SICK and the cement plant proved that in this application the GM32 has to be checked regularly only every 9 to 12 months. With the help of remote service (SICK Meeting Point Router), on-site tests were performed and a large amount of additional process data, like the CEMS concentration values at the main stack, was collected and evaluated in various aspects. During the test period it was proven that the analyzer has a stable reaction time of <20 s (fig. 5), without the need of any cleaning and smaller or larger maintenance works. The results show that with a delay of two to 17 minutes (depending on the measuring component) a CEMS alone is not sufficient for DeNOx process control.

 

Especially the NH3-value for CEMS has a huge delay in comparison to the in-situ process gas measurement (fig. 5), due to the high adsorption of ammonia on surfaces (e.g. filter and heated measurement lines). Controlling the SCR with such delayed measuring values will lead to an excessive injection of ammonia water and a faster exceedance of the NO and NH3 emission limit values at the main stack. Therefore, SICK’s GM32 in-situ measurement device has proven to be the right device for this kind of application.

 

 

Figure 5: Comparison of NH3 response times between GM32 in-situ measurement at SCR inlet and CEMS at main stack
Figure 5: Comparison of NH3 response times between GM32 in-situ measurement at SCR inlet and CEMS at main stack
Figure 5: Comparison of NH3 response times between GM32 in-situ measurement at SCR inlet and CEMS at main stack
Figure 5: Comparison of NH3 response times between GM32 in-situ measurement at SCR inlet and CEMS at main stack

 

“We are very satisfied with the performance of the GM32 gas analyzer. The GM32 project was executed fully compliant to the requirements and schedule. We are looking forward to continued good cooperation,” said Dr.-Ing. Steffen Gajewski, Plant Manager & Dipl.-Ing. Stefan Naber, Operations Manager, HeidelbergCement Geseke.

 

Remote Service and Condition Monitoring

For the future SICK and HeidelbergCement are planning to implement, SICK’s Monitoring Box in order to offer predictive maintenance services for all gas analyzers and dust measurement equipment installed on-site. The solution allows SICK to optimize maintenance, to monitor critical components or device conditions and to make remote product health inspections that recommend maintenance actions. This will enable immediate troubleshooting, which again reduces travel expenses and labor hours for service engineers. Furthermore, continuous remote condition monitoring combined with remote assistance and agreed response times assure compliance with local emission regulations, process stability and continuous production.

 

A beneficial partnership in the interest of the environment Due to the great performance of the GM32 in-situ gas analyzer and the very good cooperation between SICK and the operators of HeidelbergCement Geseke plant, another HeidelbergCement plant in Germany decided to equip its DeNOx system with two GM32 analyzers for SCR control, too. The commissioning of these devices took place in June 2020. This together with an open and intense information exchange during various environmental and process seminars have contributed to a strong partnership between SICK and HeidelbergCement.

 

 

Here is a video of another project with HeidelbergCement: