Scattered light dust measuring devices
DUSTHUNTER SP100
Your Benefits
- Ideal for thick- or double-walled ducts
- Approved according to EN 15267
- Low maintenance due to self-monitoring and contamination check
- Quick installation – no alignment required




Overview
The DUSTHUNTER SP100 is a type-approved measuring device with probe for very low to medium dust concentrations. Installation from one side makes it ideal for thick-walled or double-walled ducts. The measurement is based on the forward scattering of light. An automated check of the zero and reference point as well as a contamination check are on board in the device.
- Installation from one side
- For very low to medium dust concentrations
- Automated check of zero and reference point
- Contamination check
- Hastelloy measuring probe available for corrosive gases
- For small to medium duct diameters
- Device version for explosive zone 2/22 or 1/21
Accessories
spareparts
Applications
Service & Support
Technical overview
Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications.
- DUSTHUNTER SP100 system
DUSTHUNTER SP100 system
Measured values Scattered light intensity, dust concentration (after gravimetric comparison measurement) Performance-tested measurands Scattered light intensity Measurement principles Scattered light forward Spectral range 640 mm ... 660 mm
Laser, protection class 2, power < 1 mWMeasuring ranges Dust concentration 0 ... 5 mg/m³ / 0 ... 200 mg/m³ Higher measuring ranges on request Certified measuring ranges Dust concentration 0 ... 7.5 mg/m³ / 0 ... 10 mg/m³ / 0 ... 15 mg/m³ / 0 ... 50 mg/m³ / 0 ... 100 mg/m³ / 0 ... 200 mg/m³ / 0 ... 500 mg/m³ Response time (t90) 1 s ... 600 s
Freely adjustableAccuracy ≤ 2 %
Of measuring range full scaleProcess temperature Standard version DHSP-T2xx –40 °C ... +220 °C High temperature version DHSP-T2xx –40 °C ... +400 °C Process pressure With MCU-P control unit –50 hPa ... 10 hPa With external purge air unit –50 hPa ... 30 hPa With instrument air (provided by the customer) –100 hPa ... 100 hPa Process gas humidity Non-condensing Duct diameter ≥ 0.25 m Conformities Approved for plants requiring approval
2001/80/EC (13. BImSchV)
2000/76/EC (17. BImSchV)
27. BImSchV
TA-Luft (Prevention of Air Pollution)
EN 15267
EN 14181
MCERTS
2010/75/EU
U.S. EPA PS-11 compliant
Electrical safety CE Test functions Automatic self-test (linearity, contamination, drift, aging)
Contamination limit values: Warning at 30%; fault at 40%
Manual linearity test with reference filter
- DHSP-Txxx sender/receiver unit
DHSP-Txxx sender/receiver unit
Ambient temperature –40 °C ... +60 °C Electrical safety CE Enclosure rating IP66 Dimensions (W x H x D) Dimensions may vary. For details, see the dimensional drawings. Weight Nominal length 435 mm ≤ 6.5 kg Nominal length 735 mm ≤ 7.8 kg Nominal length 1,035 mm ≤ 9.5 kg Nominal length 1,335 mm ≤ 11 kg Power supply Voltage 24 V Voltage Supply via control unit Power consumption ≤ 4 W
- MCU-N control unit
MCU-N control unit
Description Unit to control the system components and to evaluate and output the data provided by them Ambient temperature –40 °C ... +60 °C Electrical safety CE, Protection class I Enclosure rating IP66 Analog outputs 1 output:
0/2/4 ... 20 mA, 750 Ω
Electrically isolated; two additional outputs if using I/O modules (option)Analog inputs 2 inputs:
0 ... 20 mA, 110 Ω
Not electrically isolated; two additional inputs if using I/O modules (option)Digital outputs 5 relay contacts:
48 V, 1 A
Potential-free; for status signalsDigital inputs 4 potential-free contacts Communication interface Modbus
PROFIBUS DP
Ethernet
Type of fieldbus integration TCP (via optional interface module, only one module possible per MCU) RTU RS-485 (via optional interface module, only one module possible per MCU) Via optional interface module (only one module possible per MCU) Function Connection to SOPAS ET software or OPC server Number 1 Indication LC display Status LEDs: “Power,” “Failure,” and “Maintenance request” Operation Via LC-display or software SOPAS ET Dimensions (W x H x D) 210 mm x 340 mm x 120 mm (for details see dimensional drawings) Weight ≤ 3.7 kg Power supply Voltage 90 ... 250 V Voltage Version with 24 V DC available as an option Frequency 47 ... 63 Hz Power consumption ≤ 50 W Options Interface module(s)
I/O module(s)
- MCU-P control unit
MCU-P control unit
Description Unit to control the system components and to evaluate and output the data provided by them. With integrated purge air unit. Gas flow rate ≤ 20 m³/h Ambient temperature –40 °C ... +45 °C
Intake temperatures for purge airElectrical safety CE, Protection class I Enclosure rating IP66 Analog outputs 1 output:
0/2/4 ... 20 mA, + 750 Ω
Electrically isolated; two additional outputs if using I/O modules (option)Analog inputs 2 inputs:
0 ... 20 mA, 110 Ω
Not electrically isolated; two additional inputs if using I/O modules (option)Digital outputs 5 relay contacts:
48 V, 1 A
Potential-free; for status signalsDigital inputs 4 potential-free contacts Communication interface Modbus
PROFIBUS DP
Ethernet
Type of fieldbus integration TCP (via optional interface module, only one module possible per MCU) RTU RS-485 (via optional interface module, only one module possible per MCU) Via optional interface module (only one module possible per MCU) Function Connection to SOPAS ET software or OPC server Number 1 Indication LC display Status LEDs: “Power,” “Failure,” and “Maintenance request” Operation Via LC-display or software SOPAS ET Dimensions (W x H x D) 300 mm x 455 mm x 220 mm (for details see dimensional drawings) Weight ≤ 13.5 kg Power supply Voltage 90 ... 250 V Voltage Version with 24 V DC available as an option Frequency 47 ... 63 Hz Power consumption ≤ 70 W Options Interface module(s)
I/O module(s)
- SLV4-2 purge air unit, 2BH1300, 3-ph
SLV4-2 purge air unit, 2BH1300, 3-ph
Description Unit to provide dust-free air for flushing of optical surfaces Gas flow rate 38 m³/h ... 63 m³/h
At 30 hPa counter pressure, depending on low pressure inside the filterAmbient temperature –40 °C ... +55 °C Electrical safety CE Enclosure rating IP54 Dimensions (W x H x D) 550 mm x 550 mm x 257 mm (for details see dimensional drawings) Weight 18 kg Power supply Three-phase current Δ: 200 ... 240 V, 50 Hz, 2,6 A, 400 W Y: 345 ... 415 V, 50 Hz, 1,5 A, 400 W Δ: 200 ... 275 V, 60 Hz, 2,6 A, 500 W Y: 380 ... 480 V, 60 Hz, 1,5 A, 500 W Δ: 270 ... 330 V, 50 Hz, 2,0 A, 400 W Y: 465 ... 570 V, 50 Hz, 1,16 A, 400 W Δ: 290 ... 360 V, 60 Hz, 2,1 A, 500 W Y: 500 ... 600 V, 60 Hz, 1,26 A, 500 W Δ: 230 V, 50 Hz, 2,7 A, 370 W Δ: 115 V, 60 Hz, 3,0 A, 450 W Δ: 220 ... 270 V, 50 Hz, 2,5 A, 400 W Y: 380 ... 465 V, 50 Hz, 1,45 A, 400 W Δ: 240 ... 290 V, 60 Hz, 2,6 A, 500 W Y: 415 ... 500 V, 60 Hz, 1,55 A, 500 W Auxiliary gas connections Purge air 40 mm Test functions Pressure switch (switching point –35 hPa) Integrated components 2-step air filter, type Europiclon, dust capacity 200 g