GMS800 FIDOR

Solution for continuous hydrocarbon measurements
  • At a glance
    • Standard 19” enclosure for easy integration into all industry-typical systems
    • Nearly maintenance free ejector pump delivers sample gas to the analyzer
    • Integrated catalytic converter (option) provides very clean zero gas
    • Automatic regulation and compensation of process pressure fluctuations
    • Protective filter at sample gas inlet
    • High degree of linearity (≤ 2 %) for very low and high measuring ranges
    Your Benefits
    • High availability of 99.5 %
    • Convenient remote diagnosis and operation via Ethernet using SOPAS-ET software
    • Minimum maintenance costs due to the absence of moving parts
    • Replacement of complete assemblies and modules makes repairs easy
    • Compatible with predecessor systems
    • Hydrogen as fuel gas, expensive helium is not required
    • Low operating costs, e.g., due to low hydrogen consumption (typical 30 ml/min)
    • The GMS811 FIDOR, on which the control unit (BCU) can be separated from the analyzer unit, supports convenient control and monitoring from a central control room.
    GMS800 FIDOR
    Solution for continuous hydrocarbon measurements

    A member of the GMS800 analyzer family, the GMS800 FIDOR extractive gas analyzer is ideally suited for emission monitoring according to regulatory requirements (in waste incinerators, cement and/or power plants, for example). Based on the proven flame ionization detection principle, the GMS800 FIDOR measures total hydrocarbon concentrations in gases at both trace levels and high concentration levels. The GMS800 FIDOR is available in two types: the GMS810 FIDOR with integrated control unit (BCU) and the GMS811 FIDOR with separate control unit (BCU). The housing makes integration into existing systems such as SICK's MCS100 E HW multi-component analyzer very easy.

  • Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications.

    • GMS800 FIDOR system
      Measured valuesCorg
      Performance-tested measurandsCorg
      Gas flow rate≤ 120 l/h
      Process temperature≤ +230 °C
      Process pressure–120 hPa ... 120 hPa
      Relative
      Process gas humidityNon-condensing
      Ambient temperature+5 °C ... +40 °C
      Storage temperature–20 °C ... +70 °C
      Ambient pressure900 hPa ... 1,100 hPa
      Ambient humidity≤ 95 %
      Non-condensing
      Conformities

      Approved for plants requiring approval

      2001/80/EC (13. BImSchV)

      2000/76/EC (17. BImSchV)

      27.BImSchV

      TA-Luft (Prevention of Air Pollution)

      EN 15267

      EN 14181

      MCERTS

      Electrical safetyCE
      Power supply
      Power consumptionWithout external heating
      Auxiliaries
      Instrument air≤ 1,000 l/h
      Instrument air: 4.5 ±0.5 bar
      Fuel gasTyp. 30 ml/min
      Hydrogen: 5.0 or higher, 3 ±0.2 bar
      Combustion airTyp. 250 ml/min
      Instrument air: 3 ±0.2 bar, measuring ranges below 300 ppm (500 mg/m³) require an internal
      Or external catalytic converter
      Zero gas≤ 500 l/h
      Instrument air: 3 ±0.2 bar, measuring ranges below 300 ppm (500 mg/m³) require an internal
      Or external catalytic converter
      Reference gas≤ 500 l/h
      Propane in synthetic air: 75% of measuring range full scale value
      Purge gas> 1,200 l/h
      Instrument air
      Corrective functionsAdjustment with test gases
      Test functionsExtended device diagnosis via SOPAS ET software
    • GMS810 design
      Description19″ rack enclosure with 4 rack units, for integration in cabinets
      Enclosure ratingIP40
      Dimensions (W x H x D)483 mm x 177 mm x 485 mm (for details see dimensional drawings)
      Weight≤ 17 kg
      Power supply
      Voltage90 ... 240 V
      Frequency47 ... 63 Hz
      Power consumption≤ 450 W
      Sample gas connections

      Sample gas inlet: G 1/4″, G 1/8″ double, without antikink, G 1/8″ without antikink

      Exhaust gas outlet: 12 mm straight, 1/2″ straight, 10 mm 90° bent

      Auxiliary gas connectionsAuxilary gas inlet: varies depending on type
    • GMS811 design
      Description19″ rack enclosure with 4 rack units, for usage with separate control unit (BCU), for integration in cabinets
      Enclosure ratingIP40
      Dimensions (W x H x D)483 mm x 177 mm x 352 mm (for details see dimensional drawings)
      Weight≤ 17 kg
      Power supply
      Voltage90 ... 240 V
      Frequency47 ... 63 Hz
      Power consumption≤ 450 W
      Sample gas connections

      Sample gas inlet: G 1/4″, G 1/8″ double, without antikink, G 1/8″ without antikink

      Exhaust gas outlet: G 1/4″, G 1/8″ double, without antikink, G 1/8″ without antikink

      Auxiliary gas connectionsAuxilary gas inlet: varies depending on type
    • GMS840 design
      Enclosure ratingIP 54
      Dimensions (W x H x D)522 mm x 475 mm x 478 mm (for details see dimensional drawings)
      Weight20 kg
      Power supply
      Voltage100 ... 120 V / 220 ... 240 V
      Frequency50 ... 60 Hz
      Power consumption≤ 300 W
      Sample gas connections

      Sample gas inlet

      Exhaust gas outlet: G 1/4″

    • FIDOR analyzer module
      DescriptionFlame ionization detector for measuring volatile organic components (VOC)
      Measured valuesCorg
      Performance-tested measurandsCorg
      Measurement principlesFlame ionisation detection
      Gas flow rate≤ 120 l/h
      Measuring ranges
      Corg0 ... 15 mg/m³ / 0 ... 10,000 mg/m³
      A flame arrester must be provided by the customer when measuring gas concentrations above the lower explosion limit (LEL)
      Certified measuring ranges
      Corg0 ... 15 mg/m³ / 0 ... 50 mg/m³ / 0 ... 150 mg/m³ / 0 ... 500 mg/m³
      Response time (t90)≤ 2.5 s
      Without sample gas line
      Sensitivity drift< 3 % within the maintenance interval, relative to measuring range full scale
      Zero point drift< 3 % within the maintenance interval, relative to measuring range full scale
      Detection limitCorg: 0.05 mg/m³
      Electrical safetyCE
    • BCU control unit
      DescriptionThe control unit serves as the user interface and is responsible for data processing and output as well as control and monitoring functions.
      Indication

      Status LEDs: "Power", "Maintenance" and "Failure"

      LC display

      OperationVia LC display and membrane keyboard
      Dimensions (W x H x D)375 mm x 275 mm x 66 mm (for separat BCU control unit, for details see dimensional drawings)
    • I/O module
      DescriptionClosed module with top-hat rail adapter or module for integration into enclosures
      Analog outputs4 outputs:
      0/2/4 ... 20 mA, 500 Ω
      Electrically isolated
      Analog inputs2 inputs:
      0/2/4 ... 20 mA, 500 Ω
      Not electrically isolated
      Digital inputs8 inputs:
      + 42 V
      All inlets with common reference potential
      Modbus
      Type of fieldbus integration

      TCP

      RTU RS-485

      Ethernet TCP/IP
      FunctionConnection to SOPAS ET software or OPC server

    Technical drawings

    GMS811 FIDOR
    GMS810 FIDOR
    BCU control unit
    GMS810 FIDOR and GMS811 FIDOR: Sample gas inlet on the side
    GMS810 FIDOR and GMS811 FIDOR: Sample gas inlet on the back
    GMS840 FIDOR
    GMS840 FIDOR