Contact
Feedback

TRANSIC100LP

The good nose for oxygen

  • At a glance
    • O2 transmitter based on high-performance laser spectroscopy (TDLS)
    • For use in use in explosion-hazardous areas (FM, ATEX and IECEx approvals)
    • Measurement directly in the process or extractive using a measurement gas cell (option)
    • Designed for heavy-duty industrial applications
    • Compact design and easy to operate
    • Long-term stability
    • No moving parts
    Your Benefits
    • Measures in real-time directly in the process
    • Easy installation and operation
    • Self-diagnostics with maintenance display
    • Low requirements for gas conditioning
    • Low operating costs: no consumables and no purging gas consumption
    • Rugged: reliable measurement even in contaminated gases
    TRANSIC100LP
    The good nose for oxygen

    Measuring the oxygen in the process can be quite simple. The TRANSIC100LP from SICK is the rugged transmitter that uses Tunable Diode Laser Spectroscopy (TDLS) to measure O2 reliably. A technology more common in the world of high-tech analysis, is now being used in a field instrument: at an attractive price, easy installation, a long life, low-maintenance, and suitable for use in explosion-hazardous areas.

  • Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications.

    • TRANSIC100LP system
      Measured valuesO2
      Measurement principlesDiode laser spectroscopy (TDLS)
      Measuring ranges
      O20 ... 5 Vol.-% / 0 ... 100 Vol.-%
      Response time (t90)
      ≤ 10 s
      Accuracy
      ≤ 0.2 Vol.-%
      Zero point drift
      ± 0.1 Vol.-%: per year
      Process temperature
      –20 °C ... +80 °C
      Process pressure
      800 hPa ... 1,400 hPa
      Ambient temperature
      –20 °C ... +60 °C
      Storage temperature
      –30 °C ... +80 °C
      Ex-approvals
      IECEx

      Sender/receiver unit: Ex ib IIB T4 Gb, Ex ib tb IIIC T85°C Db

      Measuring probe: Ex op is IIB T4 Ga, Ex ib tb IIIC T85°C Db

      TSA151 power supply: Ex e mb [ib] IIB T4 Gb, Ex tb [ib] IIIC T85°C Db

      ATEX

      Sender/receiver unit: II 1/2G Ex ib IIB T4 Gb, II 2D Ex ib tb IIIC T85°C Db

      Measuring probe: II 1/2G Ex op is IIB T4 Ga, II 2D Ex ib tb IIIC T85°C Db

      TSA151 power supply: II 2G Ex e mb [ib] IIB T4 Gb, II 2D Ex tb [ib] IIIC T85°C Db

      NEC/CEC (US/CA)

      Sender/receiver unit: Class I, Division 2, Group A, B, C, D T4

      Measuring probe: Class I, Division 1 + 2, Group A, B, C, D T4

      Electrical safetyCE, FM
      Enclosure rating
      IP66
      Analog outputs

      1 output active:
      0/4 ... 20 mA, 500 Ω

      1 output active:
      0/4 ... 20 mA, 200 Ω
      Only for ATEX/IECEx version TRANSIC151LP

      Digital outputs

      1 relay contact:
      + 30 V AC, 1 A / 60 V DC, 0.5 A

      1 NAMUR output:
      Only for ATEX/IECEx version TRANSIC151LP

      Serial
      RemarkNot for the ATEX/IECEx version
      Type of fieldbus integration

      RS-232

      RS-485

      FunctionProprietary service interface
      USB
      RemarkNot approved for Ex-applications
      FunctionProprietary service interface
      Dimensions (W x H x D)
      See dimensional drawings
      Weight
      2.2 kg
      Material in contact with mediaStainless steel 1.4404 (AISI 316L), Kalrez®, EPDM, PTFE, SiN, MgF2, polymer coating
      Mounting

      Flange for direct installation in the process

      Wall-mounting bracket for ambient air measurement

      Wall-mounting bracket for measurement with measuring gas cell

      Power supply
      Voltage

      24 V DC

      TRANSIC151LP: 21.6 ... 26.4 V

      For ATEX/IECEx version via TSA151 power supply, a PELV power supply is mandatory
      Current consumption

      ≤ 500 mA

      TRANSIC151LP: ≤ 240 mA

      Power consumption

      ≤ 6 W

      TRANSIC151LP: ≤ 5.2 W

      Corrective functionsAdjustment with ambient air or test gases
      Test functionsContamination check

    Technical drawings

    TRANSIC151LP, TSA151 intrinsically safe power supply
    TRANSIC100LP, weather hood for wall mounting
    TRANSIC100LP, weather hood for flange installation
    TRANSIC100LP with wall bracket for ambient measurements
    TRANSIC100LP with wall bracket and sample gas cell
    TRANSIC100LP with flange adapter PN10
    TRANSIC100LP with flange adapter for process measurements below 0.5 bar
    pin assignment
    electrical connections and pin assignment
    gas connections
TOP