Measured values | Hg |
Performance-tested measurands | Hg |
Measurement principles | Zeeman atomic absorption spectroscopy |
Gas flow rate | | | 150 l/h ... 400 l/h |
|
Measuring ranges | | Hg | 0 ... 1 ppb / 0 ... 100 ppb |
|
Certified measuring ranges | | Hg | 0 ... 10 µg/m³ / 0 ... 45 µg/m³ / 0 ... 100 µg/m³ / 0 ... 1,000 µg/m³ |
|
Sensitivity drift | | | < 3 %: within the maintenance interval, relative to measuring range full scale |
|
Zero point drift | | | < 3 %: within the maintenance interval, relative to measuring range full scale |
|
Detection limit | | | < 2 %: relative to measuring range end value |
|
Process temperature | | | ≤ +1,300 °C |
|
Process pressure | | | 850 hPa ... 1,100 hPa |
|
Ambient temperature | | MERCEM300Z | –20 °C ... +50 °C | MERCEM300Z Indoor | +5 °C ... +35 °C |
|
Storage temperature | | | –20 °C ... +40 °C |
|
Ambient pressure | | | 850 hPa ... 1,100 hPa |
|
Ambient humidity | | | ≤ 80 % Relative humidity; non-condensing |
|
Conformities | Approved for plants requiring approval 2001/80/EC (13. BImSchV) 2000/76/EC (17. BImSchV) 27. BImSchV TA-Luft (Prevention of Air Pollution) EN 15267 EN 14181 MCERTS |
Electrical safety | CE |
Enclosure rating | | MERCEM300Z | IP55 | MERCEM300Z Indoor | IP43 |
|
Analog outputs | 2 outputs: 0/4 ... 20 mA, 500 Ω Electrically isolated; additional outputs if using I/O modules (option) |
Digital outputs | 4 relay contacts: 50 V, 4 A Electrically isolated; additional outputs if using I/O modules (option) |
Digital inputs | 4 inputs: 24 V, 0.3 A Electrically isolated; additional inputs if using I/O modules (option) |
Modbus | ✔ | Type of fieldbus integration | TCP |
|
Ethernet | ✔ | Function | Connection to SOPAS ET software or OPC server |
|
|
Input | Functional keys |
Operation | Via LC-display or software SOPAS ET |
Model | Aluminum cabinet |
Dimensions (W x H x D) | | MERCE300Z with cooling device | 1,038 mm x 1,744 mm x 744 mm (for details see dimensional drawings) | MERCEM300Z Indoor | 806 mm x 2,156.5 mm x 605 mm (for details see dimensional drawings) |
|
|
Power supply | | Voltage | Depending on device version | Power consumption | Without integrated test gas generator: ≤ 2,100 VA | | With integrated test gas generator: ≤ 3,100 VA |
|
Auxiliaries | | Instrument air | ≤ 2,500 l/h 6 ... 7 bar; particle size max. 1 µm; oil content max. 0.1 mg/m³; pressure dew point max. –30 °C | Test gas | ≤ 500 l/h Max. 0.5 bar; accuracy ±2%; water content 5 ... 30 vol.-% |
|
Sample gas connections | | Sample gas inlet | Swagelok 6 mm |
|
Auxiliary gas connections | | Test gas | Swagelok 6 mm | Instrument air | Swagelok 10 mm | Exhaust gas outlet | Swagelok 10 mm |
|
Corrective functions | Internal adjustment cuvette Integrated test gas generator (option) Drift correction (QAL3) and check of the complete system |
Integrated components | Integrated conditioning of instrument air (option) Integrated test gas generator (option) |